Volumetric Modulated Arc Therapy vs. c-IMRT for the Treatment of Upper Thoracic Esophageal Cancer

  • Wu-Zhe Zhang ,

    Contributed equally to this work with: Wu-Zhe Zhang, Tian-Tian Zhai

    Affiliation Department of Radiation Oncology, Cancer Hospital of Shantou University Medical College, Guangdong, China

  • Tian-Tian Zhai ,

    Contributed equally to this work with: Wu-Zhe Zhang, Tian-Tian Zhai

    Affiliation Department of Radiation Oncology, Cancer Hospital of Shantou University Medical College, Guangdong, China

  • Jia-Yang Lu,

    Affiliation Department of Radiation Oncology, Cancer Hospital of Shantou University Medical College, Guangdong, China

  • Jian-Zhou Chen,

    Affiliation Department of Radiation Oncology, Cancer Hospital of Shantou University Medical College, Guangdong, China

  • Zhi-Jian Chen,

    Affiliation Center of Clinical Oncology, The University of Hongkong-Shenzhen Hospital 1, Shenzhen, China

  • De-Rui Li,

    Affiliation Department of Radiation Oncology, Cancer Hospital of Shantou University Medical College, Guangdong, China

  • Chuang-Zhen Chen

    stccz@139.com

    Affiliation Department of Radiation Oncology, Cancer Hospital of Shantou University Medical College, Guangdong, China

Abstract

Objective

To compare plans using volumetric-modulated arc therapy (VMAT) with conventional sliding window intensity-modulated radiation therapy (c-IMRT) to treat upper thoracic esophageal cancer (EC).

Methods

CT datasets of 11 patients with upper thoracic EC were identified. Four plans were generated for each patient: c-IMRT with 5 fields (5F) and VMAT with a single arc (1A), two arcs (2A), or three arcs (3A). The prescribed doses were 64 Gy/32 F for the primary tumor (PTV64). The dose-volume histogram data, the number of monitoring units (MUs) and the treatment time (TT) for the different plans were compared.

Results

All of the plans generated similar dose distributions for PTVs and organs at risk (OARs), except that the 2A- and 3A-VMAT plans yielded a significantly higher conformity index (CI) than the c-IMRT plan. The CI of the PTV64 was improved by increasing the number of arcs in the VMAT plans. The maximum spinal cord dose and the planning risk volume of the spinal cord dose for the two techniques were similar. The 2A- and 3A-VMAT plans yielded lower mean lung doses and heart V50 values than the c-IMRT. The V20 and V30 for the lungs in all of the VMAT plans were lower than those in the c-IMRT plan, at the expense of increasing V5, V10 and V13. The VMAT plan resulted in significant reductions in MUs and TT.

Conclusion

The 2A-VMAT plan appeared to spare the lungs from moderate-dose irradiation most effectively of all plans, at the expense of increasing the low-dose irradiation volume, and also significantly reduced the number of required MUs and the TT. The CI of the PTVs and the OARs was improved by increasing the arc-number from 1 to 2; however, no significant improvement was observed using the 3A-VMAT, except for an increase in the TT.

Introduction

Esophageal cancer (EC) is one of the most common malignant diseases in China, with a crude incidence and mortality of 19.24 and 15.39 per 100,000 residents, respectively. Treatment options for patients with non-distal metastatic thoracic esophageal carcinoma include fractionated external beam radiotherapy (EBRT) or surgery (by itself or combined with EBRT), either with or without chemotherapy. More than 60% of patients are diagnosed at locally advanced stages that cannot be completely resected. Thus, radiotherapy (RT) has become the major treatment method, especially for upper thoracic EC. However, treatment planning for RT of the upper thoracic EC is relatively challenging because these tumors are often located deep within the thorax, which is adjacent to several critical organs including the spinal cord, the lungs and the trachea. Inappropriate dosing may result in either recurrence of the disease or severe toxicity. Thus, it is important to investigate novel radiation delivery techniques to improve the dose coverage of this tumor and spare normal tissues.

Conventional intensity-modulated radiation therapy (c-IMRT) can be used to provide superior coverage of the planning target volume (PTV) compared to three-dimensional conformal radiation therapy (3DCRT)[13]. Many clinical studies have yielded good dosimetric results and patient treatment outcomes using IMRT. VMAT is a novel form of IMRT in which radiation treatment is delivered during gantry rotation with dynamic multi-leaf collimator (MLC) motion, variable dose rates (DR) and gantry speed modulation. Many studies have shown that VMAT can produce dosimetrically equivalent plans to IMRT for centrally located cancers, such as prostate cancer, cervical cancer and head and neck cancers[46].VMAT also requires a shorter treatment time (TT) and fewer monitoring units (MUs) than IMRT, thereby overcoming the primary drawbacks of IMRT. A shorter treatment time may make patients feel more comfortable and reduce motion errors during treatment delivery. Decreasing the amount of MUs used could reduce undesirable irradiation of normal tissue and minimize the risk of secondary cancer.

In a study by Benthuysen, the use of VMAT yielded equivalent tumor coverage and protection for organs at risk (OARs) and PTV as using IMRT, while reducing the required MUs and TTs in distal esophageal cancer[7]. However, none of the previously published reports, except the aforementioned study by Liyin[8], have addressed these issues for upper thoracic EC.

The objective of this pilot study was to evaluate the feasibility of using VMAT for thoracic EC and to formulate an optimal plan design for this technique by comparing tumor coverage, OAR sparing, MUs and TTs for VMAT and c-IMRT plans. The effect of arc number on the dose distribution in the VMAT plans was also analyzed.

Materials and Methods

Patients

CT datasets were identified for 11 patients (8 males and 3 females), who had pathologically confirmed upper thoracic esophageal squamous cell carcinomas and underwent radical radiotherapy at Shantou University Cancer Hospital in 2008. This retrospective study was approved by the Ethics Committee at Shantou University Medical College Tumor Hospital. All of the participants provide their written informed consent to participate in this study. The lengths of the tumor lesions ranged from 4.5 to 10.5 cm, with a median of 6.0 cm and a mean of 6.3 cm. The volumes of the esophageal tumors were 5.9–50.95 cm3, with a median of 21.0 cm3 and a mean of 25.1 cm3. The tumor characteristics are summarized in Table 1.

Simulation

All of the patients were immobilized with a head and upper thoracic thermoplastic mask in a supine position with their arms alongside their bodies. All of the CT datasets were acquired using a helical CT scanner (Philips Brilliance CT Big Bore Oncology Configuration, Cleveland, OH, USA). The CT images were taken at a 5-mm thickness throughout the entire thorax and neck that extending to 10 cm beyond the borders of the tumor. The data were transferred to the Eclipse 10.0 treatment planning system (Varian Medical Systems, Palo Alto, CA, USA) according to the DICOM 3.0 protocol.

Delineation of target volumes and organs at risk

The gross tumor volume (GTV) and clinical target volumes (CTVs) were contoured by senior physicians. The GTV, including the esophageal cancer and involved positive regional lymph nodes, was defined using CT images, barium swallow fluoroscopy and endoscopy. Two CTVs were contoured to cover potential microscopic diseases. The CTV64 was delineated with 5–10 mm radial and 20 mm longitudinal margins with respect to the GTV. The CTV54 was derived from the CTV64 plus the lymph node draining area in the upper mediastinum and bilateral supraclavicular region. Irrelevant bony structure and lung tissue were excluded from the CTVs. The PTV was created by adding a 5-mm margin to the relative CTV. The following OARs were considered: the lungs, the heart and the spinal cord with its planning risk volume (PRV-spinal cord, spinal cord plus a 5-mm margin).

Planning techniques and objectives

All of the treatment plans were generated to deliver 64 Gy and 54 Gy to the PTV64 and PTV54, respectively, in 32 fractions using the Eclipse 10.0 treatment planning system (Varian Medical Systems, Palo Alto, CA, USA) with a 6-MV photon beam from a Varian Truebeam linear accelerator (Varian Medical Systems, Palo Alto, CA, USA), which was equipped with an MLC with 120 leaves (with a spatial resolution of 5 mm at the isocenter for the central 20 cm and 10 mm for the outer 20 cm). Both of the plans were optimized at a maximum DR of 600 MU/min. Continuous gantry motion, dose-rate variation, and MLC motion were approximated by optimizing individual beams at 2° gantry angle increments. The dose calculation was performed using the anisotropic analytical algorithm (AAA, version 8.6.02) using a 2.5-mm grid. The planning objective for the PTV64 was at least 95% of the PTV volume that received 100% of the prescribed dose, with Dmin > 60 Gy, Dmax < 70 Gy and V105% < 5%. The following dose constraints were defined for the OARs: spinal cord, Dmax < 40 Gy; PRV-Spinal cord, Dmax < 45 Gy; heart, V30 < 30%; and lung, V20 < 25% and V30 < 20%. Fulfilling the objectives for the PTV was a higher priority than fulfilling those for the OARs.To meet the prescribed and the OAR-limited, planning goals, dose-volume constraints such as volume and weight were fine-tuned for all of the plans during inverse planning using the direct machine parameter optimization method. To ensure consistency of the planning techniques, all of the treatment plans were devised by physicists with over 8 years of clinical experience in IMRT, and over 3-years in VMAT planning.The planning requirement and techniques for planners were also aligned by training with standard protocols, and procedures in the department.

c-IMRT plans

The c-IMRT plans were created using five coplanar fields (5F; gantry angles: 0°, 72°, 144°, 216° and 288°). The MLC leaf sequences of the plans were generated using a dynamic fixed field sliding window IMRT delivery technique.

VMAT plans

Coplanar arcs with full gantry rotation and opposite rotation (clockwise or counter-clockwise) were used for each plan and were optimized simultaneously: a single arc for 1A, two arcs for 2A and three arcs for 3A (clockwise, counter-clockwise or clockwise, respectively). The field size and the collimator rotation were determined using an automatic tool from Eclipse to encompass the PTV. Each individual arc was limited to a sequence of 177 control points; therefore, two and three coplanar arcs were used to increase the modulation factor during optimization to simultaneously achieve increase the target homogeneity and reduce the involvement of the OARs.

After treatment planning was completed, the plan was normalized to cover 95% of the PTV with 100% of the prescribed dose. In the present study, the constraints and priority factors in the c-IMRT and VMAT plans were modified to optimize the results by adjusting the parameters as a function of the dose-volume histogram (DVH) data for each patient.

Evaluation tools

The lans pwere evaluated using the DVH data. Dx was the specific dose that was computed for a fraction of a target or an organ volume, and Vx was the volume that was irradiated above a designated dose. The mean dose (Dmean) and the maximum dose (Dmax) for the PTV were analyzed, where the CI of PTV64 was defined as (1)

(1)

The homogeneity index (HI) of the PTV was computed as in (2): (2) (i.e.the difference between the dose covering 5% and 95% of the PTV). The DVH parameters for the OARs (the spinal cord, the PRV-spinal cord, the heart and the lungs) were calculated and compared using a t-test.

Statistical analyses

The Statistical Package for Social Science (SPSS, Version 19.0, Chicago, IL, USA) was used to perform the statistical analyses. A normal distribution test showed that all of the data were normally distributed. The data for the different plans were reported as the average ± standard deviation (X±S) and compared using a paired two-tailed t-test. Differences in the data were considered to be statistically significant for P < 0.05.

Results

Target coverage, conformity and dose homogeneity

All of the VMAT and c-IMRT plans were clinically acceptable. The VMAT and c-IMRT results are listed in Table 2. The DVH of the PTV for one patient is shown in Fig. 1. The 2A and 3A VMAT plans generated a significantly better CI for PTV54 than the c-IMRT plan. There was no significant difference between the 2A and 3A plans, except that the V105 was lower for the 3A plan (P < 0.05), thus favoring the 3A plan. Additionally, the 3A plan had a better HI for PTV64 than the c-IMRT plan (P < 0.05), whereas similar HI values were obtained for the 2A and c-IMRT plans. The HI value for the 1A plan was slightly inferior to that of the c-IMRT plan (P < 0.05). There were minor, but statistically significant, differences in the V93, V95 and V105 values of PTV54 between the VMAT and c-IMRT plans. The conformity and homogeneity of the VMAT plans improved as more arcs were used.

thumbnail
Fig 1. Comparison of dose-volume histograms (DVHs) for four plans.

https://doi.org/10.1371/journal.pone.0121385.g001

thumbnail
Table 2. Dosimetric results for planning target volume (PTV) ().

https://doi.org/10.1371/journal.pone.0121385.t002

OARs

The dosimetric results for the spinal cord, PRV-spinal cord, heart and lungs are summarized in Table 3. The DVH for the OAR in one patient is shown in Fig. 1. The Dmax values for the spinal cord and the PRV-spinal cord were similar for the 2A and c-IMRT plans, whereas the 3A plan showed a slight and significant trend toward better results than the c-IMRT plan (P < 0.05). Only the 1A plan achieved a higher Dmax than the c-IMRT plan (P < 0.05). The VMAT plan generated a significantly lower V20 and V30 for the lungs than the c-IMRT plan (P < 0.05) as V5, V10 and V13 were incrementally increased. Likewise, the Dmean of the lungs and the heart V50 was lower for the 2A and 3A plan than for the 1A or c-IMRT plans. No significant difference was detected between the 2A and 3A plans or the 1A and c-IMRT plans.

thumbnail
Table 3. Dosimetric results for organs at risk (OARs) ().

https://doi.org/10.1371/journal.pone.0121385.t003

MUs and TTs

The delivery efficiency (in terms of the number of MUs and the TT) is shown in Table 4.

thumbnail
Table 4. Dosimetric results for number of monitoring units (MUs) and TTs ().

https://doi.org/10.1371/journal.pone.0121385.t004

The number of MUs for the 1A (331±18), 2A (345±16) and 3A (347±14) plans were reduced by 50%, 48% and 48%, respectively, over that of the c-IMRT plan (663±71). We calculated TT, and the result was IMRT(183±12 sec)、1A(68±4 sec)、2A(137±7 sec) and 3A(222±15 sec). All of the plans were also performed in the QA mode to record the TTs. Compared with the c-IMRT plan, there were substantial reductions of 67% and 21% in the TTs for the 1A and 2A plans (62±1 sec and 149±1 sec vs. 189±15 sec), whereas the TT of the 3A plan increased by 26% (239±1 sec).Plans 1A and 2A had a faster delivery than c-IMRT.

Discussion

Brahme et al. developed a novel method in which the order of conventional treatment planning was reversed: the optimum incident beam dose distributions was determined from the desired dose distribution in the target volume in 1988[9]. Since then, c-IMRT has gradually become the preferred planning method for many cancers. A study has shown that the PTV coverage for a c-IMRT plan with 5F was comparable with those for 7F or 9F c-IMRT plans for upper thoracic EC. Fu et al.[10] found that the conformity of the c-IMRT plans did not improve further by increasing the number of fields beyond five. Wang et al.[11] reported that both 5- and 7F c-IMRT plans were optimal methods for upper thoracic EC. Therefore, we designed the c-IMRT plans with five fields in this study.

Over the past few years, it has been shown that VMAT, a novel form of IMRT, is dosimetrically superior to c-IMRT[5,1213]. The most appropriate number of arcs used in the plans with respect to different disease sites remains a subject of debate. However, using more arcs in VMAT enables the superior modulation of factors during optimization by taking advantage of the independent optimization, the unrelated sequence of the MLC shape, the gantry speed and the dose rate combinations. Thus, increasing the number of arcs in a VMAT plan further optimizes the dose distribution in PTV in terms of conformity and homogeneity. However, VMATs with a larger number of arcs generally require more time for treatment delivery and quality assurance. Therefore, the appropriate arc number of VMATs remains to be determined.

The conformity and homogeneity of the dose distribution in the PTV are determined by the target volumes, the delivery equipment and the radiotherapy technology. Several studies have shown that a 1A VMAT plan may be dosimetrically less favorable than a fixed beam IMRT plan[1416].

Our results showed that the 2A and 3A plans generated a superior CI for the PTV to the c-IMRT plan (0.88 and 0.89 vs. 0.86, P = 0.046,0.008), whereas there was no significant difference between the 1A VMAT and c-IMRT plans, except that the c-IMRT plan had a better HI than the 1A VMAT plan. Further analysis indicated that increasing the arc number of the VMAT plan improved the CI (0.88 and 0.89 for the 2A and 3A plans vs. 0.85 in the 1A plan) and the HI (1.07 and 1.06 in the 2A and 3A plans vs. 1.08 in the 1A plan) of the PTV. However, no significant difference was found between the 2A and 3A plans.

Significant advances have been made in radiotherapy delivery techniques over the past decades. These developments have been primarily driven by the need to reduce the radiation dose and the volume of normal tissues and organs, thereby minimizing the risks of toxicity and morbidity. Studies have shown that VMAT can deliver radiation treatment with equivalent or better OAR sparing than IMRT[8,17].

The spinal cord was given the highest priority for sparing among the OARs, and the Dmax of the spinal cord has normally been limited to <45 Gy by NCCN guidelines. However, recently published reports have suggested a Dmax for the spinal cord that is <40 Gy when RT is combined with chemotherapy[1819]. This goal was met by the 2A, 3A and c-IMRT plans, whereas the Dmax of the spinal cord in the 1A VMAT plan was slightly higher than 40 Gy: thus, the 1A VMAT plan may not be suitable for chemoradiotherapy of EC in terms of protecting the spinal cord.

The lung is another important OAR that should be considered during treatment planning. Our data showed that the dose constraints of the lungs were satisfied by all four plans. Moreover, compared to the c-IMRT plan, the 2A and 3A VMAT plans significantly reduced the lung volume that was irradiated with a moderate dose. Thus, the 2A and 3A VMAT plans have the potential to reduce the incidence of radiation pneumonitis. However, although the volume of lungs that was irradiated with a moderate dose (20 Gy or 30 Gy) was reduced using VMAT, a greater lung volume received a low dose (≤13 Gy) irradiation, which could compromise the prior advantage. It is unclear to what extent this consideration could affect the overall incidence of radiation pneumonitis. A study by Wang et al. [2021] found that V5 was the only significant factor associated with treatment-related pneumonitis. A recent study conducted by MD Anderson showed that when the V5 of the lungs was limited to <65%, the risk of radiation pneumonitis was 24% for grade 2 and 12% for grade 3, and the related mortality rate was 1%. Thus, it was recommended that V5 be limited to within 65%[22]. In the current study, we observed only slight increases in V5 (1A 48.0, 2A 48.1, 3A 48.1 and 5F 47.3), V10 (1A 41.5, 2A 41.4, 3A 41.5 and 5F 39.6) and V13 (1A 35.9, 2A 35.3, 3A 35.5 and 5F 32.8) using the VMAT plans. More importantly, V5 in all of the plans was far below 65%. Thus, increasing low-dose irradiation of the lungs using VMAT should not compromise the advantage this plan offers in sparing OARs. For the heart, the reduction trend of heart parameters (V30, V40 and V50) was similar between 2A-VMAT and c-IMRT, and no difference in Dmean or V20 was found between VMAT and c-IMRT.

Our results also showed that the number of MUs required using the 1A (331±18), 2A (345±16) and 3A (347±14) plans were reduced by 50%, 48% and 48% compared to the number of MUs required for the c-IMRT plan (663±71), which was consistent with the results of a study by Clivio[20]. The significant increase in the number of MUs required in the IMRT plan has become a growing concern. An increase in the number of MUs may increase undesirable irradiation of normal tissues via the scattered dose, leading to an elevated risk of secondary cancers after treatment[21]. By contrast, the VMAT plan requires many fewer MUs because this plan is performed simultaneously with gantry rotation and a dynamic MLC adaptation to the target volume, instead of the sliding window technique that is used in c-IMRT. Thus, the application of VMAT may potentially reduce secondary cancer after radiation treatment.

Finally, VMAT also has a faster delivery than c-IMRT. In our study, the TTs for the 1A (62±1 sec) and 2A (149±1 sec) plans were reduced by 67% and 21%, respectively, compared to the c-IMRT plan (189±15 sec) although the TT of plan 3A (239±1sec) was increased by 26%. These results were consistent with those from a study on the fractional irradiation of malignant glioma in which the VMAT was 3.3 times as fast as IMRT[23]. Thus, the shorter treatment time and fewer MUs that are required for the VMAT plan compared with the c-IMRT plan could result in the VMAT plan becoming an important tool in image guided radiotherapy and hyper-fractionated accelerated radiotherapy[24].

A recent study[25]demonstrates that, using VMAT(two arcs)for patients with upper esophageal carcinoma, significantly reduces the delivery time and the dose to the lungs compared with IMRT. Our study’s conclusion on this point is similar. Furthermore, we compared VMAT plans with different number of arcs, and found that VMAT with two arcs is a good choice of radiotherapy techniques for upper esophageal carcinoma.

In conclusion, the 2A-VMAT plan appeared to spare the lungs from moderate-dose irradiation more than any of the other plans, at the expense of increasing the volume of low-dose irradiation, and also significantly reduced the number of required MUs and the TT over those of the c-IMRT plan. The dose conformity of the PTVs and OARs were improved when the arc number of the VMAT plan was increased from one to two; however, no further significant improvement was observed using the 3A-VMAT plan, except for an increased TT.

Author Contributions

Conceived and designed the experiments: CZC. Performed the experiments: WZZ TTZ. Analyzed the data: WZZ TTZ. Contributed reagents/materials/analysis tools: WZZ TTZ JYL JZC ZJC DRL. Wrote the paper: TTZ.

References

  1. 1. Wu Q, Manning M, Schmidt-Ullrich R, Mohan R. The potential for sparing of parotids and escalation of biologically effective dose with intensity-modulated radiation treatments of head and neck cancers: a treatment design study. Int J Radiat Oncol Biol Phys, 2000,46:195–205. pmid:10656393
  2. 2. Wu VW, Kwong DL, Sham JS. Target dose conformity in 3-dimensional conformal radiotherapy and intensity modulated radiotherapy. Radiother Oncol,2004,71:201–206. pmid:15110454
  3. 3. Chandra A, Guerrero TM, Liu HH, Tucker SL, Liao Z, Wang X, et al.Feasibility of using intensity-mudulated radiotherapy to improve lung sparing in treatment planning for distal esophageal cancer. Radiother Oncol, 2005,77:247–253. pmid:16298001
  4. 4. Wolff D, Stieler F, Welzel G, Lorenz F, Abo-Madyan , Sabine M, et al. Volumetric modulated arc therapy (VMAT) vs.serial tomotherapy, step-and-shoot IMRT and 3D-conformal RT for treatment of prostate cancer. Radiother Oncol,2009,93(2):226–233. pmid:19765846
  5. 5. Wagner D, Christiansen H, Wolff H, Vorwerk H. Radiotherapy of malignant gliomas:Comparison of volumetric single arc technique(RapidArc),dynamic intensity-modulated technique and 3D conformal technique. Radiother Oncol,2009,93(3):593–596. pmid:19897266
  6. 6. Bertelsen A, Hansen CR, Johansen J, Brink C. Single Arc Volumetric Modulated Arc Therapy of head and neck cancer.Radiother Oncol,2010,95(2):142–148. pmid:20188427
  7. 7. Liam Van Benthuysen , Lee Hales, Matthew B. Volumetric modulated arc therapy VS. IMRT for the treatment of distal esophageal cancer. Medical Dosimetry, 2011,36,(4):404–409. pmid:21377864
  8. 8. Yin L, Wu H, Gong J, Geng JH, Jiang F, Shi AH, et al. Volumetric-modulated arc therapy vs c-IMRT in esophageal cancer: A treatment planning comparison. World J Gastroenterol,2012,18(37):5266–5275. pmid:23066322
  9. 9. Brahme A. Optimization of stationary and moving beam radiation therapy techniques. Radiother Oncol,1988,12:129–140. pmid:3406458
  10. 10. Fu WH, Wang LH, Zhou MZ, Dai JR, Hu YM, Zhao LJ. Comparison of conformal and intensity-modulated techniques for simultaneous integrated boost radiotherapy of upper esophageal carcinoma.World J Gastroenterol,2004,10(8):1098–1102. pmid:15069706
  11. 11. Wang J, Zhu S-C, Han C, Chi Z-F, Liu D, et al. Comparison of dose distributions between conformal and intensity-modulated radiotherapy for the upper esophageal carcinoma. Int J Radiat Oncol Biol Phys, 2008,17(40): 275–279.
  12. 12. Cozzi L, Dinshaw KA, Shrivastava SK, Mahantshetty U, Engineer R, Deshpande DD, et al. A treatment planning study com-paring volumetric arc modulationwith RapidArc and fixed field IMRT for cervix uteri radiotherapy.Radiother Onco2008,l89: 180–191
  13. 13. Mayo CS, Ding L, Addesa A, Kadish S, Fitzgerald TJ, Mos-er R. Initial experience with volumetric IMRT (RapidArc) for intracranial stereotactic radiosurgery. Int J Radiat Oncol Biol Phys, 2010,78: 1457–1466 pmid:20207494
  14. 14. Lee TF, Ting HM, Chao PJ, Fang FM. Dual arc volumetric-modulated arc radiotherapy (VMAT) of nasopharyngeal carcinomas: a simultaneous integrated boost treatment plan comparison with intensity-modulated radiotherapies and single arc VMAT.Clin Oncol,2012,24(3):196–207. pmid:21752615
  15. 15. Guckenberger M, Richter A, Krieger T, Wilbert J, Baier K, Flentje M, et al. Is a single arc sufficient in volumetric-modulated arc therapy (VMAT) for complex-shaped target volumes?. Radiother Oncol 2009,(2),93:259–265. pmid:19748146
  16. 16. Zhang DD, Huang SM, Deng XW, Zhang GS, Huang XY, Sun WZ, et al. Comparison and evaluation of VMAT and IMRT for the treatment of initial treated nasopharyngealcarcinoma.Chinese Journal of Radiation Oncology2012,(21), 4:364–368.
  17. 17. Van Benthuysen L, Hales L, Podgorsak MB.Volumetric modulated arc therapy vs. IMRT for the treatment of distal esophageal cancer. Med Dosim. 2011, 36(4):404–409. pmid:21377864
  18. 18. Burmeister BH, Walpole ET, D'Arcy N, Burmeister EA, Cox S, Thomson DB, et al. A phase II trial of chemoradiation therapy with weekly oxaliplatin and protracted infusion of 5-fluorouracil for esophageal cancer.Invest New Drugs. 2009,Jun;27(3):275–279. pmid:18841327
  19. 19. Pera M, Gallego R, Montagut C, Martín-Richard M, Iglesias M, Conill C, et al. Phase II trial of preoperative chemoradiotherapy with oxaliplatin, cisplatin, and 5-FU in locally advancedesophageal and gastric cancer.Ann Oncol.2012, Mar;23(3):664–670. pmid:21652581
  20. 20. Graham MV, Purdy JA, Emami B, Harms W, Bosch W, Lockett MA, et al. Clinical dose volume histogram analysis for pneumonitis after 3D treatment for non small cell lung cancer(NSCLC). Int J Radiat,1999,45(2):323. pmid:10487552
  21. 21. Shi AH, Zhu GY, Wu H, Yu R, Li FH, Xu B. Analysis of clinical and dosimetric factors associated with treatment-related pneumonitis(TRP) in patients with non-small-cell lung cancer(NSCLC) treated with concurrent chemotherapy and three-dimensional conformal radiotherapy (3D-CRT).Int J Radiat Oncol Biol Phys,2006,66(5):1399–1407. pmid:16997503
  22. 22. Jiang ZQ, Yang K, Komaki R, Wei X, Tucker SL, Zhu Y, et al. Long-Term Clinical Outcome of Intensity-Modulated Radiotherapy for Inoperable Non-samll-Cell Lung Cancer:The MD Anderson Experience. Int J Radiat Oncol Biol Phys,2012,83(1):332–339. pmid:22079735
  23. 23. Hall EJ, Wuu CS.Radiation-induced second cancers: the impact of 3D-CRT and IMRT.Int J Radiat Oncol Biol Phys. 2003, May 1;56(1):83–8. pmid:12694826
  24. 24. Amendola BE, Amendola M, Perez N, Iglesias A, Wu X. Volumetric-modulated arc therapy with RapidArc: An evaluation of treatment delivery efficiency.Rep Pract Oncol Radiother.2013, Aug 17;18(6):383–386. pmid:24416583
  25. 25. Ma P, Wang X, Xu Y, Dai J, Wang L. Applying the technique of volume-modulated arc radiotherapy to upper esophageal carcinoma. J Appl Clin Med Phys2014,15(3)221–228.