Correction: Reliability and Validity of a Point-of-Care Sural Nerve Conduction Device for Identification of Diabetic Neuropathy

  • The PLOS ONE Staff
  • Article
  • Metrics
  • Comments
  • Media Coverage

There is an error in the penultimate sentence of the Results section of the Abstract. The correct sentence is: A SNAP of ≤6 µV had 88% sensitivity and 94% specificity for identifying age-and height-standardized reference NCS values, while a SNCV of ≤48 m/s had 94% sensitivity and 82% specificity.”

Reference

  1. 1. Lee JA, Halpern EM, Lovblom LE, Yeung E, Bril V, et al. (2014) Reliability and Validity of a Point-of-Care Sural Nerve Conduction Device for Identification of Diabetic Neuropathy. PLoS ONE 9(1): e86515