A New Approach to Detect Congestive Heart Failure Using Short-Term Heart Rate Variability Measures

  • Guanzheng Liu ,

    Contributed equally to this work with: Guanzheng Liu, Lei Wang

    Affiliation School of Engineering, Sun Yat-sen University, Guangzhou, China

  • Lei Wang ,

    Contributed equally to this work with: Guanzheng Liu, Lei Wang

    Affiliation Shenzhen Institutes of Advanced Technology, the Chinese Academy of Sciences, Shenzhen, China

  • Qian Wang,

    Affiliation School of Engineering, Sun Yat-sen University, Guangzhou, China

  • GuangMin Zhou,

    Affiliation School of Engineering, Sun Yat-sen University, Guangzhou, China

  • Ying Wang,

    Affiliation School of Engineering, Sun Yat-sen University, Guangzhou, China

  • Qing Jiang

    jqing@mail.sysu.edu.cn

    Affiliation School of Engineering, Sun Yat-sen University, Guangzhou, China

Abstract

Heart rate variability (HRV) analysis has quantified the functioning of the autonomic regulation of the heart and heart's ability to respond. However, majority of studies on HRV report several differences between patients with congestive heart failure (CHF) and healthy subjects, such as time-domain, frequency domain and nonlinear HRV measures. In the paper, we mainly presented a new approach to detect congestive heart failure (CHF) based on combination support vector machine (SVM) and three nonstandard heart rate variability (HRV) measures (e.g. SUM_TD, SUM_FD and SUM_IE). The CHF classification model was presented by using SVM classifier with the combination SUM_TD and SUM_FD. In the analysis performed, we found that the CHF classification algorithm could obtain the best performance with the CHF classification accuracy, sensitivity and specificity of 100%, 100%, 100%, respectively.

Introduction

Autonomic dysfunction is a typical feature of chronic heart failure and is associated with severity of disease and prognosis in chronic heart failure (CHF) [1]. As a simple noninvasive technology, heart rate variability (HRV) analysis provides reliable information on autonomic modulation of heart rate, and it has been a valuable tool to understand psychopathological mechanisms of heart failure. Indeed, the significant difference of heart rate variability between patients with chronic heart failure and healthy people was widely reported in previous studies [2]. However, CHF is asymptomatic in its first stages. Therefore, early assessment of CHF severity is crucial to avoid the condition worsening and contribute to decrease medical cost.

The heart rate variability (HRV) measures have been mainly studied for the prognosis of the disease, in particular, as predictor of the risk of mortality. Several studies addressed the relationship between HRV and CHF [3][4]. For example, Nolan et al. (1998) prospectively showed that SDNN (standard deviation of all RR intervals) was a strong independent prognostic tool in CHF patients [5]. The low-frequency spectral component LF (ranging between 0.03 and 0.15 Hz) decreased in CHF patients with advanced disease and was related to the progression of the heart failure [6]. Guzzetti et al (2000) also suggested information content present in non-linear analysis of HRV in CHF patients has prognostic relevance independently from the time domain and spectral analysis of HRV [7]. However, HRV parameters were inevitably disturbed by its spontaneous fluctuation [8], respiration [9], motion artifacts [10], and ambulatory health-monitoring application and so on. As a result, some guide evidences that HRV analysis shouldn't form the primary basis for CHF assessment, because of its sensitivity and specificity.

Conventional, the New York Heart Association (NYHA) classification, which is a symptomatic functional scale, is one of the most widespread assessment methods of the severity of CHF [11]. As an objective evaluation, the comprehensive 2-D echocardiogram coupled with Doppler flow was also widely used for assessment of CHF [12]. However, an interesting question is whether HRV analysis may improve both sensitivity and specificity of ECG examination, thus providing a robust independent tool for CHF assessment.

Recently, a small number of attentions are being paid to CHF assessment based on heart rate variability (HRV) measurement. Asyali et al. (2003) adopted conventional time and frequency parameters of HRV and a Bayesian classifier to discriminate CHF disease with sensitivity and specificity rate of 81.82% and 98.08%, respectively [13]. Isler et al (2007) investigated the discrimination power of combining wavelet entropy and conventional HRV parameters by using the genetic algorithms and k-nearest neighbor classifier. The best performance was obtained with sensitivity and specificity rate of 100.00% and 94.74%, respectively [14]. Pecchia et al (2011) presented the CART algorithm and tree-like decision classifier to achieve CHF classification based on the conventional time-/frequency-parameters, including of two nonstandard feature, e.g. ΔAVNN (average of all RR intervals) and ΔLF/HF [15]. The result demonstrated the sensitivity and specificity rate were 89.7% and 100%, respectively. Although these studies reached interesting results, they all use superabundant HRV parameters, which could affect the discrimination sensitivity or increase the complex of classifier for the daily activity of clinicians.

In the paper, we presented a CHF patients' classification algorithm based on three new nonstandard HRV measures and support vector machine (SVM).

Methods and Experiments

Data

The dataset used in this work was obtained from an online and widely-used database, i.e. MIT/BIH database [16]. This study was approved by the Institutional Review Boards of Beth Israel Deaconess Medical Center (Boston, MA) and the Massachusetts Institute of Technology (Cambridge, MA). All subjects provided informed written consent and all data were de-identified. Patient records/information was anonymized and de-identified prior to analysis. Two RR interval database were chosen, including: 1) Thirty healthy subjects with a mean age of 49.33±19.77 years (range 40–72 years); 2) Seventeen patients with CHF in NYHA I–III with a mean age of 60.88±50.01 years (range 51–71 years). The data of healthy subjects was retrieved from the normal sinus rhythm RR interval database. The data for the CHF group was retrieved from the congestive heart failure RR interval database. The RR interval records are provided with beat annotations obtained by automated analysis with manual review and correction. According to the reference [9], the 5 minutes RR intervals were extracted from the beginning stage in the database. The original ECG records were digitalized at 128 samples per second.

CHF Classification Algorithm

The CHF classification model are comprised of short HRV feature computing, features processing, and Modeling based on support vector machine (SVM) as figure 1. Furthermore, the feature processing involves a three-part process of feature-selection, feature-normalization and feature-combination.

thumbnail
Figure 1. The CHF classification algorithm based on support vector machine.

https://doi.org/10.1371/journal.pone.0093399.g001

entropy-based measures of short term HRV.

We extracted 5-min RR interval time series (RRITS) excerpts from the 24-h records and obtained 12 HRV features as the table 1.

The ApEn and SampEn were computed based on the following equation (15).

First It is assumed that RR time series is the data set {RR (k), k = 1, 2…N} with length N. then, the phase space is reconstructed by choosing two parameters: the embedding dimension (m) and the delay () [18]. The delay () is 1 beat and embedding dimension (m) is 2 in the paper. The (N-m+1) templates are composed as follows:(1)

The recurrence matrix and distance between matrices are calculated as equation (2) and (3):(2)(3)

Where is constant, (e.g. one standard deviation from RR time series).

Thus, the Approximate Entropy is defined as:(4)

Then, for each :(5)

Since approximate entropy indicates more similarity than it's truly one for finite time series. To reduce the bias caused by it matching, sample entropy (SampEn) was developed to quantify heart rate variability [19]. The equations (2) and (4) are adjusted:(6)

Then the SampEn is computed as following equation:(7)

Three short-term nonstandard HRV measures.

Feature processing in a high-dimensional data space is often used to decrease the computational cost and improve the accuracy during the classification process [20]. The feature processing procedure is included of feature-selection, feature-normalization and feature- combination in our study as following:

Step-1: In contrast to the feature-extraction techniques, feature selection is used to identify the variables that do not contribute to the classification process. In the paper, we assessed all the front 12 HRV measures (also known as features) by using the statistical method (with SPSS software, vision 14 (SPSS Inc, Chicago, IL, USA)). First, mean and standard deviations is used to evaluate the mean absolute error between the healthy group and the CHF group. Then, two datasets are compared by using two sample t-tests for each HRV feature. The significance level chosen was α = 0.1. The bigger the significance parameter (*p) is, the smaller the contribution is. At last, if the significance parameter (*p) is greater than 0.1, the HRV feature was rejected. Table 2 demonstrated that three HRV features (e.g. rMSSD, VHF and LF/HF) were deleted with the significance parameter (*p>0.1).

thumbnail
Table 2. HEART RATE VARIABILITY (HRV) MEASURES BETWEEN CHF DISEASES AND HEALTHY PEOPLE.

https://doi.org/10.1371/journal.pone.0093399.t002

Step-2: Feature normalization was used to independently normalize each feature component to the [0 1] range [21]. Feature normalization could reduce the differences among HRV measures. It assumes that the feature vector X {X (i), i = 1, 2…N}. The normalization's feature (Y) was obtained as following equation:(8)(9)(10)

Step-3: our study supposed that the same type of feature (e.g. time domain, frequency domain and information entropy) has extremely similar difference between CHF group and healthy group. Thus, feature-combination could further reduce the dimensionality and computational cost. Three new nonstandard features are combined of the nine features (e.g. SDNN, pNN50%, CVrr, VLF, LF, HF, TP, ApEn and SampEn) as the following equation:

The nonstandard time-domain feature was computing as following:(11)

The nonstandard frequency-domain feature was computing as following:(12)

The nonstandard non-line feature was computing as following:(13)

CHF classification algorithm based on support vector machine (SVM) and three nonstandard features.

As one of the most popular classifiers, Support vector machine (SVM) aims to find a hyperplane that can separate two classes of data with maximal margin [22]. The n-dimensional feature vector is x and the bias b belongs to {−1, 1}. The decision function f(x) was computed as following:(14)

The weight vector w and bias b representing a separating hyperplane. The SVM finds an optimal separating margin by solving the following optimization task:

Subject to:(15)

Where, the parameters were the slack variables, the constant parameter C is obtained by a cross-validation method. The polynomial kernel function was chosen as following equation:(16)

K nearest-neighbor (KNN) classifier for CHF classification

To evaluate our algorithm (section II_B) performance, a k-nearest neighbor classifier [14] was also investigated in the paper. By comparison among previous CHF classification methods, we found the K nearest-neighbor (KNN) classifier [14] has the best performance. Moreover, the different inputs form time domain, frequency domain and non-line features were considered to obtain optimal CHF classification performance. The KNN classifier with the combination of different features was also studied in terms of SEN and AUC.

Performance measurements

To measure the performance of CHF classification algorithm, we use the confusion matrices [23]. From these matrices, we compute the widely used measures reported in table 3 for binary classification in order to enable the comparison of our method with others. The software program Mathworks Matlab version 7 (R2007b) (The MathWorks, Natick, MA, USA) is used for data processing in the paper. We did the statistical tests (with SPSS software, vision 14 (SPSS Inc, Chicago, IL, USA)) for HRV measures choice. Where two datasets are compared, we perform two sample t-tests for each individual method. The significance level is chosen as α = 0.1.

Results

The performance of three nonstandard features

The fig. 2 demonstrates the difference of three nonstandard features (SUM_TD, SUM_FD and SUM_IE) between patients with CHF and healthy people. In patients with CHF, the SUM_TD (307.47±129.0 versus 75.60±44.86), is significant higher than one of healthy people with the significant level (*p<0.0001). By comparison with all time-domain features, the combinational feature (SUM_TD) has bigger difference. In patients with CHF, the SUM_FD (18.29±6.60 versus 23.78±8.79), the SUM_IE (0.68±0.56 versus 2.37±0.77) are significant lower than those of healthy people with the significant parameter (*p) of 0.019, 0.0001, respectively. In comparison with all frequency, the SUM_FD has more small significant parameter (*p). The SUM_IE has also bigger difference than single information entropy feature (e.g. ApEn and SampEn).

thumbnail
Figure 2. Three nonstandard features between patients with CHF and healthy people; Mean ±one standard deviation was plotted.

SUM_TD was a nonstandard time domain feature; SUM_FD was a nonstandard frequency domain feature; SUM_IE was a nonstandard non-line feature.

https://doi.org/10.1371/journal.pone.0093399.g002

Thus, three nonstandard features significantly magnify the difference between patients with CHF and healthy people. For example, the frequency feature (SUM_FD) has more significantly diffidence (*p<0.019) that all frequency features (e.g. VLF, LF, HF, LF/HF, TP and VHF). Thus, the feature-processing obviously magnifies the difference, and reduces the dimension and the computational cost of the CHF disease classification model based on SVM classifier.

CHF classification performances among different feature combination

The study found that single nonstandard feature was still unable to separate patients with CHF from healthy people with 100% accuracy. In comparison, the SUM_IE is best classification parameter with the best accuracy of 95.74%. The SUM_FD has worst classification accuracy of 59.57%.

The figure 3 demonstrates four kinds of CHF classification model based different input feature vectors. The blue-dotted lines (fig. 3) show the linear separating curves for three kinds of input feature vectors (e.g. SUM_TD, SUM_FD; SUM_TD, SUM_IE; and SUM_IE, SUM_FD). The results suggest that the input feature vector with SUM_TD and SUM_FD was linear completely separable. Obviously, if the three nonstandard features are all chosen as input feature vector, the CHF classification model should also be linear separable. The input feature vector with SUM_TD and SUM_IE aren't linear separable with two linear inseparable samples. The input feature vector with SUM_FD and SUM_TD is also linear inseparable with five linear inseparable samples. Moreover, three black curves in the fig. 3 are the separating curves of SVM based on polynomial kernel. And the result implies the SVM classifier' effect of the polynomial kernel is better than one of linear kernel.

thumbnail
Figure 3. The SVM classifier from different input feature vectors.

SUM_TD was a nonstandard time domain feature; SUM_FD was a nonstandard frequency domain feature; SUM_IE was a nonstandard non-line feature.

https://doi.org/10.1371/journal.pone.0093399.g003

CHF classification algorithm

By comparison with K nearest-neighbor (KNN) classifier [14] (as table 4), our algorithm could obtain better performance of CHF classification. The results demonstrated that the KNN classifier with all 12 features has the lowest discrimination power in all terms (ACC, PER, SEN, SPE and AUC). The performance of different CHF classification models based is demonstrated as table 5. The result suggests that the optional classification models could achieve accuracy (ACC), precision (PRE), sensitivity (SEN), specificity (SPE), area under the curve (AUC) of 100.00%, 100.00%,100.00%, 100%, 100%, 100.00% as following input feature vector: (a) SUM_TD, SUM_FD and SUM_IE; (b) SUM_TD and SUM_FD; (c) all 12 features. However, considered to the computation cost, the CHF classification model with the input features (SUM_TD and SUM_FD) is chosen in the study. Besides, other input feature vectors (e.g. SUM_TD and SUM_IE; SUM_IE and SUM_FD) also obtain preferably performance with the classification accuracy of 97.98%.

thumbnail
Table 4. THE K-NEAREST-NEIGHBOR CLASSIFER FOR CHF CLASSIFICATION.

https://doi.org/10.1371/journal.pone.0093399.t004

thumbnail
Table 5. THE PERFORMANCE OF CHF classification MODEL BASED ON SVM.

https://doi.org/10.1371/journal.pone.0093399.t005

The support vector is computed by SVM classifier:(17)

Where the parameters (p) is: [0.0001 0.0208 9.3036]; the parameter (x) is SUM_TD; the parameter (y) means output value from the model.

Thus, the CHF classification model is obtained by SVM:(18)

If Z<0 represents CHF patients.

Discussion and Conclusion

In this study, we aim to investigate the CHF classification power from three nonstandard short-term HRV features (e.g. SUM_TD, SUM_FD and SUM_IE) based on SVM classifier. The CHF classification model (equation 18) we presented could achieve to discriminate CHF patients from healthy people with the accuracy values of 100%. Moreover, the three nonstandard HRV features enhance the difference between healthy people and patients with CHF, and contribute to CHF classification.

Bilchick and Berger reported that a depressed SDNN (<70 ms and <30 ms, respectively, for long-term and 5-min records) is significantly associated with increased mortality [24]. In our study, the depressed SUM_IE (<1.18) is significantly associated with CHF diseases with the significant parameter (*p<0.001). The SUM_FD had worst CHF classification performance with the accuracy value of 59.57%. However, when the input features chose SUM_TD and SUM_IE, the classification accuracy achieved to 100%. The high complementarity between nonstandard frequency-domain feature (SUM_TD) and nonstandard time-domain feature (SUM_IE) have contributed to improve the CHF classification's performance.

Compared to the other studies in table 6, we found that the long-term HRV measures haven't contributed to improve the CHF classification's performance. The KNN classifier from the paper has a bit worse performance than the KNN classifier form Isler et al. The possible reason is that the study form Isler et al chose more HRV measures (e.g. Wavelet entropy and Poincare plot). The study form Isler et al has more sample numbers with the 29 patients and 54 healthy subjects. Obviously, the SVM classifier from the paper has best performance for CHF classification.

thumbnail
Table 6. CHF CLASSIFICATION PERFORMANCE ASSESSMENT OF DIFFERENT CLASSIFIERS.

https://doi.org/10.1371/journal.pone.0093399.t006

In the study, we could draw some conclusions:

  1. Three short-term nonstandard HRV measures (e.g. SUM_TD, SUM_FD and SUM_IE) we presented are suited to CHF classification. In fact, the nonstandard HRV features are better than conventional standard HRV measures in CHF classification.
  2. The SVM classifier with the combination of nonstandard time-domain and nonstandard frequency-domain have the highest discrimination power with the classification accuracy, precision, sensitivity, specificity of 100%, 100%, 100%, 100%, respectively. In fact, the SVM classifier is superior to KNN classifier. The most serious problem for KNN classifier is the feature chooses and combination.
  3. The study also found that the short-term HRV measures could substitute for long-term HRV measures in the CHF classification. The CHF classification performance with long-term HRV measures is not significantly above one with short-term HRV measures.

Finally, the proposed CHF classification algorithm meets all our requirements because it is fully noninvasive, low-cost and high accuracy method that provides an objective classification for classification of CHF.

Author Contributions

Conceived and designed the experiments: GL QJ. Performed the experiments: QW GZ. Analyzed the data: GL LW. Contributed reagents/materials/analysis tools: GL LW QJ. Wrote the paper: GL. Experimental sorting: YW.

References

  1. 1. Packer M (1992) The neurohormonal hypothesis: a theory to explain the mechanism of disease progression in heart failure. J Am Coll Cardiol 20: 248–254.
  2. 2. Nolan J, Batin PD, Andrews R, Lindsay SJ, Brooksby P, et al. (1998) Prospective study of heart rate variability and mortality in chronic heart failure: results of the United Kingdom heart failure evaluation and assessment of risk trial (UK-heart). Circulation 98: 1510–1516.
  3. 3. Moore RKG, Groves D, Kearney MT, Eckberg DL, Callahan TS, et al. (2004) HRV spectral power and mortality in chronic heart failure (CHF): 5 year results of the UK heart study. Heart 90: A6.
  4. 4. Guzzetti S, Magatelli E, Borroni E, Mezzetti S (2001) Heart rate variability in chronic heart failure. Autonomic Neuroscience: Basic and Clinical 90: 102–105.
  5. 5. Nolan J, Batin PD, Andrews R, Lindsay SJ, Brooksby P, et al. (1998) Prospective study of heart rate variability and mortality in chronic heart failure. Circulation 98 (15) 1510–1516.
  6. 6. Guzzetti S, Cogliati C, Turiel M, Crema C, Lombardi F, et al. (1995) Sympathetic predominance followed by functional denervation in the progression of chronic heart failure. European Heart Journal 16 (8) 1100–1107.
  7. 7. Guzzetti S, Mezzetti S, Magatelli R, Porta A, De Angelis G, et al. (2000) Linear and non-linear 24 h heart rate variability in chronic heart failure. Autonomic neuroscience basic & clinical 86 (1–2) 114–119.
  8. 8. Hu J, Gao JB, Tung WW, Cao YH (2010) Multiscale Analysis of Heart Rate Variability: A Comparison of Different Complexity Measures. Annals of biomedical engineering 38 (3) 854–864.
  9. 9. Liu GZ, Huang BY, Wang L (2011) A Wearable Respiratory Biofeedback System Based on Generalized Body Sensor Network. Telemedicine and e-health 17 (5) 348–357.
  10. 10. Liu GZ, Guo YW, Zhu QS, Huang BY, Wang L (2011) Estimation of Respiration Rate from Three-Dimensional Acceleration Data Based on Body Sensor Network. Telemedicine and e-health 17 (9) 705–711.
  11. 11. Dorgin M (1994) Nomenclature and Criteria for classification of Diseases of the Heart and Great Vessels. New York: Little Brown and Company.
  12. 12. Jessup M, Abraham WT, Casey DE, Theodore GG, Mariell J, et al. (2009) Focused update: ACCF/AHA guidelines for the classification and management of heart failure in adults a report of the American college of cardiology foundation/American heart association task force on practice guidelines. Circulation 119 (14) 1977–2016.
  13. 13. Asyali MH (2003) Discrimination power of long-term heart rate variability measures. Proc 25th Annu Int Conf IEEE Eng Med Biol Soc 200–203.
  14. 14. Isler Y, Kuntalp M (2007) Combining classical HRV indices with wavelet entropy measures improves to performance in diagnosing congestive heart failure. Comput Biol Med 37 (10) 1502–1510.
  15. 15. Pecchia L, Melillo P, Sansone M, Bracale M (2011) Discrimination power of short-term heart rate variability measures for CHF assessment. IEEE Transactions on Information Technology in Biomedicine 15 (1) 40–46.
  16. 16. Available: http://www.physionet.org/cgi-bin/ATM (Accessed 2013 Dec 12.)
  17. 17. Malik M (1996) Standards of measurement, physiological interpretation, and clinical use. Circulation 93: 1043–1065.
  18. 18. Miguel AGG, Mireya FC, Juan RC (2009) Errors in the Estimation of Approximate Entropy and Other Recurrence-Plot-Derived Indices Due to the Finite Resolution of RR Time Series. IEEE Transactions on Biomedical Engineering 56 (2) 345–351.
  19. 19. Haitham MAA, Alan VS (2007) Use of Sample Entropy Approach to Study Heart Rate Variability in Obstructive Sleep Apnea Syndrome. IEEE Transactions on Biomedical Engineering 54 (10) 1900–1904.
  20. 20. Chang CI, Wang S (2006) Constrained band selection for hyperspectral imagery. IEEE Trans. Geosci. Remote Sens 44 (6) 1575–1585.
  21. 21. Aksoy S, Haralick RM (2001) Feature normalization and likelihood-based similarity measures for image retrieval. Pattern Recognition letters 22 (5) 563–582.
  22. 22. Xue H, Chen SC, Yang Q (2011) Structural regularized support vector machine: a framework for structural large margin classifier. IEEE Transactions on Neural Networks 22 (4) 573–587.
  23. 23. Provost F, Kohavi R (1998) Guest editor's introduction: on applied research in machine learning. Mach, Learn 30 (2) 127–132.
  24. 24. Bilchick KC, Berger RD (2006) Heart rate variability. J. Cardiovasc. Electropysiol 17 (6) 691–694.