Evaluation of Microbubbles as Contrast Agents for Ultrasonography and Magnetic Resonance Imaging

  • Ling Li,

    Affiliation Jiangsu Key Laboratory of Molecular Imaging and Functional Imaging, Department of Radiology, Zhong-Da Hospital, Medical School of Southeast University, Nanjing, China

  • Qiang Wei,

    Affiliation Department of Ultrasound, the Second Affiliated Hospital, Southeast University, Nanjing, China

  • Hong-Bo Li,

    Affiliation Department of Ultrasound, the Second Affiliated Hospital, Southeast University, Nanjing, China

  • Song Wen,

    Affiliation Jiangsu Key Laboratory of Molecular Imaging and Functional Imaging, Department of Radiology, Zhong-Da Hospital, Medical School of Southeast University, Nanjing, China

  • Gao-Jun Teng

    gjteng@seu.edu.cn

    Affiliation Jiangsu Key Laboratory of Molecular Imaging and Functional Imaging, Department of Radiology, Zhong-Da Hospital, Medical School of Southeast University, Nanjing, China

Abstract

Background

Microbubbles (MBs) can serve as an ultrasound contrast agent, and has the potential for magnetic resonance imaging (MRI). Due to the relatively low effect of MBs on MRI, it is necessary to develop new MBs that are more suitable for MRI. In this study, we evaluate the properties of SonoVue® and custom-made Fe3O4-nanoparticle-embedded microbubbles (Fe3O4-MBs) in terms of contrast agents for ultrsonography (US) and MRI.

Methodology/Principal Findings

A total of 20 HepG2 subcutaneous-tumor-bearing nude mice were randomly assigned to 2 groups (i.e., n = 10 mice each group), one for US test and the other for MRI test. Within each group, two tests were performed for each mouse. The contrast agent for the first test is SonoVue®, and the second is Fe3O4-MBs. US was performed using a TechnosMPX US system (Esaote, Italy) with a contrast-tuned imaging (CnTI™) mode. MRI was performed using a 7.0T Micro-MRI (PharmaScan, Bruker Biospin GmbH, Germany) with an EPI-T2* sequence. The data of signal-to-noise ratio (SNR) from the region-of-interest of each US and MR image was calculated by ImageJ (National Institute of Health, USA). In group 1, enhancement of SonoVue® was significantly higher than Fe3O4-MBs on US (P<0.001). In group 2, negative enhancement of Fe3O4-MBs was significantly higher than SonoVue® on MRI (P<0.001). The time to peak showed no significant differences between US and MRI, both of which used the same MBs (P>0.05). The SNR analysis of the enhancement process reveals a strong negative correlation in both cases (i.e., SonoVue® r = −0.733, Fe3O4-MBs r = −0.903, with P<0.05).

Conclusions

It might be important to change the Fe3O4-MBs' shell structure and/or the imagining strategy of US to improve the imaging quality of Fe3O4-MBs on US. As an intriguing prospect that can be detected by US and MRI, MBs are worthy of further study.

Introduction

Angiogenesis is a determinant of tumor growth, invasion, and metastasis [1]. To detect new tumor microvessels, modern medical imaging modalities are widely used. As non-invasive imaging tools, ultrasonography (US) and magnetic resonance imaging (MRI) are becoming very popular.

With few exceptions, ultrasound contrast agents (UCAs) are microbubbles (MBs), which are often 1–7 µm in diameter and are used primarily as blood-pool markers [2]. Since the first description of enhanced reflections of ultrasound in 1969 [3], UCAs have developed rapidly and the existing MBs usually have an inert gas core (sulfur hexafluoride or perfluorocarbon gases) and a stable shell (denatured albumin, surfactants, or phospholipids). Typical MRI contrast agents include gadolinium chelate, manganese chelate, and iron compounds. However, in 1991, Moseley indicated that gas-filled MBs could also be used as a unique MR contrast agent [4]. Recent theoretical and phantom studies had further demonstrated this [5][9]. The principle behind their use in MRI was the gas-liquid interface or the pressure-induced microbubble size changing, which induced large local magnetic susceptibility differences.

The current research about MBs and MRI mainly falls into two categories: (1) using MBs as an MR contract agent based on the magnetic susceptibility of MBs [10][11], and (2) using MBs as a medium in MRI based on the biological effects produced by MBs' cavitation and sonoporation characteristics [12][14]. The first category of research has received little attention in the literature, mainly due to the relatively low effect of MBs on MRI. On the other hand, some researchers found that the potential application of MBs as a unique intravascular susceptibility contrast agent for MRI has not been fully studied. Along this line, references [4], [10] have focused on the feasibility study in vivo of MRI with existing MBs, and references [11], [15][19] mainly focused on developing new MBs that are more suitable for MRI.

In the present study, we are interested in determining the ability of microbubbles as contrast agents for ultrasonography and magnetic resonance imaging.

Results

In vitro US experiments

The signal strength was 92.08±7.45, 56.53±4.86 for SonoVue® and Fe3O4-MBs, respectively. There was a significant difference (P<0.001). When the imaging strategy changed from contrast-tuned imaging (CnTI™) to the Flash mode, the SonoVue® microbubbles broke and the enhanced signal generated by the microbubbles changed to anecho (the signal strength was 8.62±3.45, P<0.001, Fig. 1A). However, under the same imaging strategy change, the signal of Fe3O4-MBs tube had little change as few microbubbles broke in this case (the signal strength was 50.54±6.37, P>0.05, Fig. 1B).

thumbnail
Figure 1. In vitro US experiments.

Images of SonoVue® and Fe3O4-MBs under the mode of CnTI™ and Flash: (A) In the first half of the tube, in CnTI® mode, SonoVue® showed high echo; in the second half, in Flash mode, the SonoVue® microbubbles broke and the enhanced signal generated by the microbubbles changed to anecho (white arrow). (B) Under the same imaging strategy change, the signal of Fe3O4-MBs tube had little change (white arrow).

https://doi.org/10.1371/journal.pone.0034644.g001

In vitro MRI experiments

Figure 2A includes MR images of different concentration in SonoVue® and Fe3O4-MBs suspension phantoms. The signal strength from Fe3O4-MBs was lower than SonoVue®. Figure 2B shows the dependency of the SNR on different SonoVue® volume fractions. Figure 2C shows the dependency of the SNR on different Fe3O4 -MBs volume fractions. An approximately linear relationship was observed independently (r = −0.982 for SonoVue®, r = −0.929 for Fe3O4-MBs, with P<0.05).

thumbnail
Figure 2. In vitro MRI experiments.

The in vitro MR images of different volume fraction of SonoVue® and Fe3O4-MBs: (A). 1: sodium chloride solution (0.9% w/v), 2–4: SonoVue® with different volume fraction (1.75%, 3.5%, 7.0%). 5–7: Fe3O4-MBs with different volume fraction (1.75%, 3.5%, 7.0%). (B) an approximately linear relationship was observed in SonoVue® (r = −0.982, P<0.05) between SNR and microbubble volume fraction. (C) an approximately linear relationship was observed in Fe3O4-MBs (r = −0.929, P<0.05) between SNR and microbubble volume fraction.

https://doi.org/10.1371/journal.pone.0034644.g002

In vivo US imaging

Generally, about 21 days after subcutaneous injection of tumor cells, the tumor maximum diameter was close to 0.77±0.08 cm. MB contrast enhancement was observed in all 10 mice by US and all 10 by MRI.

Figure 3 illustrates the images typically observed by US with SonoVue® and Fe3O4-MBs. Figure 3A shows the gray-scale image of the tumor. Under CnTI™ mode just before the MBs injection, signals from stationary tissues were suppressed and only high amplitude signals were visualized (Fig. 3B). After the MB suspension injection, the signal of the tumor enhanced. Figures 3C and 3D demonstrate the maximum contrast after injection SonoVue® and Fe3O4-MBs, respectively.

thumbnail
Figure 3. Representative US findings with SonoVue® and Fe3O4-MBs.

Images from the same mouse: (A) gray-scale image of the tumor with a size 0.79 mm×0.34 mm, (B) under CnTI™ mode just before microbubble injection, signals from tumor was suppressed (white arrow), (C) maximum contrast enhancement US tumor image after SonoVue® injection (white arrow), and (D) maximum contrast enhancement US tumor image after Fe3O4-MBs injection (white arrow).

https://doi.org/10.1371/journal.pone.0034644.g003

Moreover, Figure 4 illustrates the time-course signal changes induced by SonoVue® (Figure 4A) and Fe3O4-MBs (Figure 4B) injection from the same region-of-interest (ROI) and the same tumor. The average value of enhanced signal observed was 26.14±10.95 and 8.52±5.83 for SonoVue® and Fe3O4-MBs during the entire imaging process, respectively, and there was a significant difference (P<0.001).

thumbnail
Figure 4. SNR time courses of the same mouse tumor in US by using SonoVue® and Fe3O4-MBs.

Figure A shows the SNR time course in ROI during SonoVue® injection and Figure B shows the SNR time course when using Fe3O4-MBs injection.

https://doi.org/10.1371/journal.pone.0034644.g004

In vivo MR imaging

Typical time courses of EPI-T2* images show the similar change trends as US after MBs administration. Figure 5 illustrates the mouse tumor images typically observed during SonoVue® and Fe3O4-MBs injection. Figure 5A shows the anatomy of superficial tumor. Figure 5B illustrates pre-injection EPI-T2* images. The post-injection images of SonoVue® and Fe3O4-MBs correspond to the lowest SNR point in Figures 5C and 5D.

thumbnail
Figure 5. Representative MRI findings with SonoVue® and Fe3O4-MBs.

Images from the same mouse during microbubble injection: (A) anatomical image showing the superficial tumor (white arrow), (B) pre-injection EPI-T2* MRI tumor image (white arrow), (C) maximum contrast enhancement of the tumor after SonoVue® injection (white arrow), and (D) maximum contrast enhancement of the tumor after Fe3O4-MBs injection (white arrow).

https://doi.org/10.1371/journal.pone.0034644.g005

Moreover, Figure 6 illustrates the SNR time-course signal changes induced by SonoVue® (Figure 6A) and Fe3O4-MBs (Figure 6B) injection from the same ROI and the same tumor. The average signal strength was 51.57±5.01 and 43.80±8.38 for SonoVue® and Fe3O4-MBs respectively, and there was a significant difference (P<0.001).

thumbnail
Figure 6. SNR time courses of MRI from SonoVue® and Fe3O4-MBs.

Figure A shows the SNR time course in ROI during SonoVue® injection and Figure B shows the SNR time course when using Fe3O4-MBs.

https://doi.org/10.1371/journal.pone.0034644.g006

Correlation between the US and MRI

Time-to-peak values were 33.60±1.58 s and 34.20±1.55 s in SonoVue® for US and MRI, respectively, and there was no significant difference (P = 0.402). For Fe3O4-MBs, the corresponding values are 26.50±1.27 s and 23.80±1.69 s for US and MRI, respectively, and there was no significant difference either (P = 0.404).

Figure 7A denotes the negative correlation of the tumor signal strength change between US and MRI when both using SonoVue®. The corresponding correlation is r = −0.733 with p<0.05. Figure 7B shows a similar negative correlation when we use Fe3O4-MBs in US and MRI instead. The corresponding correlation in this case is r = −0.903 with p<0.05. The above SNR analyses revealed a strong and significant relationship between the two modalities when using the same microbubble contrast agent.

thumbnail
Figure 7. Correlation between the US and MRI.

Figure A shows the signal strength correlation between US and MRI in 600 seconds of enhancement when using SonoVue® (r = −0.733, P<0.05). Figure B shows the signal strength correlation between US and MRI in 600 seconds of enhancement when using Fe3O4-MBs (r = −0.903, P<0.05).

https://doi.org/10.1371/journal.pone.0034644.g007

Discussion

Despite the enhancement in different ways, the two typical SNR time courses showed the similar trend. The tumors were enhanced rapidly first and washed out gradually. The time of dynamic enhancement was only a few minutes after injection because of the limited lifetime of MBs in vivo.

In group 1, Fe3O4-MBs had a lower effect on US than SonoVue®. The reasons are mainly (1) the Fe3O4-MBs is a relatively new material and has a hard shell, their acoustic properties (e.g., acoustic impedance) are somewhat different from phospholipid-stabilized MBs under the same acoustic energy, so the intensity of second harmonic is relatively weak. (2) Most of the medical equipment today has been designed and tested based on the properties of SonoVue®. Without changing the imaging strategies, it is difficult to achieve good imaging properties with the new Fe3O4-shelled microbubbles. We have conducted many tests to improve the imaging quality on the existing medical devices. However, methods such as using higher frequencies or Doppler based destructive imaging cannot achieve the desirable effects yet. We are working with several Chinese equipment manufactures on improving the imaging qualities, hopefully obtaining some positive results soon.

In group 2, however, SonoVue® had a lower effect on MRI than Fe3O4-MBs, and the effect of SonoVue® had reached its limit according to the approaches described for improving microbubbles for MRI, which included increasing the radius and volume fraction of MBs, using inert gas core, et al [8], [10]. First, as the size SonoVue® ranges from 0.1 µm to 1.1 µm (with a mean diameter is 2.5 micrometer according to manual), thus the diameter should not be increased significantly or they will not pass through the pulmonary circulation. Second, the microbubble concentration used in our study was 0.1 mL of the ∼3.5% volume fraction, which was far more than the common maximum clinical dosage (e.g., 0.08 mL/kg of the 0.8% volume fraction according to the product description) [10]. Third, the gas core is already sulfur hexafluoride. Since an inert gas has a high molecular weight and low solubility, it can cause the largest effect on transverse relaxation.

Comparing with SonoVue®, the most important change in the custom-made MBs was that some Fe3O4 nanoparticles had been embedded into a double-polymer shell. Fe3O4 are T2 agents as MBs, and they can enhance the magnetic sensitivity for the MBs [11], [18]. As the thick shell [20], the susceptibility of US is reduced. Changing the characteristics of the shell and/or the imagining strategy on US may be the most important factor in improving the effectiveness for US. After all, a combination of multiple modalities can offer synergistic advantages over any modality alone [21].

In a closely related paper [22], the authors reported an early contrast study of tumor perfusion using US and MRI. They used MBs as the contrast agent for US and gadopentetate dimeglumine as the contract agent for MRI. And found that using MBs in US can achieve the same effect as using gadopentetate dimeglumine in MR imaging. As an intriguing prospect that can be detected by US and MRI, MBs are worthy of further study.

Materials and Methods

In vitro US experiments

A microbubble phantom study was performed (L.L., Q.W., H.B.L.) to measure SNR of SonoVue® and Fe3O4-MBs suspension using a TechnosMPX US system (Esaote, Italy) with a linear-array transducer (LA532E, 7.5 MHz for fundamental gray-scale imaging and 2.5 MHz for CnTI™). Fe3O4-MBs was kindly provided by Dr. Fang Yang [18], which had double polymer shells with 86.47 µg/mL Fe3O4 nanoparticles in it and N2 gas core and the mean diameter was 3.98 µm. The samples were well mixed and placed in silica gel tubes of 1.0 cm in diameter. They were diluted to 3.5% volume fractions with sodium chloride solution (0.9% w/v) and put in a de-gassed water tank. Two experiments were conducted, one under CnTI™ imaging and the other under Flash imaging. The CnTI™ parameters were as follows: the gain was 105, the depth was 31 mm, and the mechanical index (MI) was 0.089, while in the Flash mode the MI was 0.5.

In vitro MRI experiments

A microbubble phantom study was performed (L.L.,S.W.) to measure SNR of Fe3O4-MBs and SonoVue® suspension using a 7.0T Micro-MRI (PhamaScan, Bruker, Germany). The samples were mixed and placed in Eppendorf tubes of 1.0 cm in diameter. They were diluted to 7.0%, 3.5%, and 1.75% volume fractions with sodium chloride solution (0.9% w/v) respectively. Every phantom was scanned three times. The imaging parameters for T2-weighted fast spin echo were set as repetition time (TR) = 3000 ms, echo time (TE) = 60 ms, number of excitations (NEX) = 1.

Microbubble preparation for in vivo experiments

The first contrast agent used was SonoVue®. The second contrast agent used was Fe3O4-MBs. For each group, the dosages are all 0.1 mL of microbubble suspension with a volume fraction of about 3.5%. Before each injection, re-suspension must be performed.

Cell culture and animal model

All animal procedures were performed in accordance with the approval and guidelines of the Institutional Animal Care and Use Committee (IACUC) of the Medical School of Southeast University (approval ID: SYXK-2007.2121).

Cells from the human hepatocellular liver carcinoma cell line HepG2 (Keygen Biotech. Co., Ltd, Nanjing, China) were grown in RPMI 1640 with 10% fetal bovine serum (FBS) in a 5% CO2 humidified atmosphere at 37°C. Tumors were established in 20 healthy BALB/c-nu mice (5–6 weeks old, 20–25 g in weight) in random order by subcutaneous injection of a suspension of 2×107 HepG2 cells in 0.2 mL of phosphate-buffered saline (PBS) in the right leg. Tumors were allowed to grow until the greatest diameter of the tumors was close to 0.8 cm.

Animal preparation

A total of 20 HepG2 subcutaneous-tumor-bearing nude mice were randomly assigned to 2 groups (i.e., n = 10 mice each group), one for the US test, and the other for the MRI test. Within each group, two tests were performed for each mouse. The contrast agent for the first test is SonoVue®, and the second is Fe3O4-MBs. For US, all mice were kept anesthetized with intraperitoneal injection of 10% chloral hydrate (3 mL/kg). For MRI, mice were anesthetized with 1.0% isoflurane via a nose core with respiratory monitoring. A 30-G needle was inserted into a tail vein by intravenous injection of the bubbles with a 40-cm long PE-10 tube. The dose of MBs was 0.1 mL and 0.2 mL of saline flush was administered immediately after MBs injected.

In vivo US experiments

Before each examination, each mouse was fixed in a left lateral position on a warm pad to maintain the body temperature. As the tumor was just under the skin, about a 2-mm deep US gel was placed between the transducer surface and the skin to ensure the tumor was imaged clearly by US.

US imaging was performed (L.L., Q.W., H.B.L.) using a TechnosMPX US system under CnTI™ mode. The CnTI™ parameters were maintained during all examinations: the gain was 105, the depth was 31 mm, the MI was 0.089 and the scan time was 1 s. When the optical imaging plane was obtained (the tumor's largest transverse cross section), the transducer was maintained with a mechanical fixer. MB suspension was injected about 5 min after the CnTI™ was triggered and images were recorded digitally on a hard disk for 15 min for off-line analysis. The data of SNR from the ROI of each US image was calculated by ImageJ (a software by National Institute of Health, USA). The time-course of the corresponding parameter SNR was measured.

For each mouse, SonoVue® check was done first. After the sufficient clearance of the MBs, Fe3O4-MBs suspension was injected.

In vivo MRI experiments

MRI was performed (L.L., S.W.) using a 7.0T Micro-MRI with a 38-mm volume coil. Each mouse was placed in the prone position and dynamic susceptibility imaging was performed using an EPI-T2* sequence with the respiratory gating control. To obtain better images and shorten the scan time as much as possible, parameters were adjusted as follows: TR = 1000 ms; TE = 30 ms; field of view (FOV) = 4×4 cm; FA = 90°; NEX = 1; scan time = 1 s; slice thickness = 1 mm. Anatomical images were acquired under another protocol using the following parameters: TR = 2500 ms; TE = 33 ms; FOV = 4×4 cm; FA = 180°; NEX = 1; scan time = 1 min20 s; slice thickness = 1 mm. MB suspension was injected about 5 min after the start of the dynamic imaging and the total scanning time was 15 min. All images were recorded digitally on a hard disk for off-line analysis. ROI analyses were conducted as described for US and the SNR time course was also measured.

For each mouse, SonoVue® check was done first then was the Fe3O4-MBs.

Statistical analyses

Statistical evaluation was performed using SPSS software (ver.13.0; SPSS Inc., Chicago, IL, USA). The numerical results were expressed as means±SD. Results were analyzed statistically using the paired-samples t test (for comparing the average signal strengths of using different MBs in US or MRI) and independent-samples t test (for comparing the time-to-peak average signal strengths of using the same MBs in US and MRI) for effectiveness of different MBs and modalities. Pearson correlation coefficient of SNR is for US and MRI.

Acknowledgments

We thank Prof. Hua Zhuang (Department of Ultrasound, the Second Affiliated Hospital, Southeast University, Nanjing, China) and Prof. Chi-Min Zhang (Department of Ultrasound, Zhong-Da Hospital, Southeast University, Nanjing, China) for their helpful discussions.

Author Contributions

Conceived and designed the experiments: LL GJT. Performed the experiments: LL QW HBL SW. Analyzed the data: LL QW HBL SW. Wrote the paper: LL GJT.

References

  1. 1. Folkman J (2006) Angiogenesis. Annu Rev Med 57: 1–18.
  2. 2. Cosgrove D (2006) Ultrasound contrast agents: An overview. Eur J Radiol 60: 324–330.
  3. 3. Voigt JU (2009) Ultrasound molecular imaging. Methods 48: 92–97.
  4. 4. Wong KK, Huang I, Kim YR, Tang H, Yang ES, et al. (2004) In vivo study of microbubbles as an MR susceptibility contrast agent. Magn Reson Med 52: 445–452.
  5. 5. Alexander AL, McCreery TT, Barrette TR, Gmitro AF, Unger EC (1996) Microbubbles as novel pressure-sensitive MR contrast agents. Magn Reson Med 35: 801–806.
  6. 6. Dharmakumar R, Plewes DB, Wright GA (2002) On the parameters affecting the sensitivity of MR measures of pressure with microbubbles. Magn Reson Med 47: 264–273.
  7. 7. Dharmakumar R, Plewes DB, Wright GA (2005) A novel microbubble construct for intracardiac or intravascular MR manometry: a theoretical study. Phys Med Biol 50: 4745–4762.
  8. 8. Ueguchi T, Tanaka Y, Hamada S, Kawamoto R, Ogata Y, et al. (2006) Air microbubbles as MR susceptibility contrast agent at 1.5 Tesla. Magn Reson Med Sci 5: 147–150.
  9. 9. Francois DG, Hugues BC, Armel D (2008) Signal decay due to susceptibility-induced intravoxel dephasing on multiple air-filled cylinders: MRI simulations and experiments. Magn Reson Mater Phy 21: 261–271.
  10. 10. Cheung JS, Chow AM, Guo H, Wu EX (2009) Microbubbles as a novel contrast agent for brain MRI. Neuroimage 46: 658–664.
  11. 11. Chow AM, Chan KWY, Cheung JS, Wu EX (2010) Enhancement of gas-filled microbubble R2* by iron oxide nanoparticles for MRI. Magn Reson Med 63: 224–229.
  12. 12. McDannold N, Zhang Y, Vykhodtseva N (2011) Blood-brain barrier disruption and vascular damage induced by ultrasound bursts combined with microbubbles can be influenced by choice of anesthesia protocol. Ultrasound Med Biol 37: 1259–1270.
  13. 13. Tung YS, Vlachos F, Choi JJ, Deffieux T, Selert K, et al. (2010) In vivo transcranial cavitation threshold detection during ultrasound-induced blood-brain barrier opening in mice. Phys Med Biol 55: 6141–6155.
  14. 14. Vlachos F, Tung YS, Konofagou EE (2010) Permeability assessment of the focused ultrasound-induced blood-brain barrier opening using dynamic contrast-enhanced MRI. Phys Med Biol 55: 5451–5466.
  15. 15. Liu Z, Lammers T, Ehling J, Fokong S, Bornemann J, et al. (2011) Iron oxide nanoparticle-containing microbubble composites as contrast agents for MR and ultrasound dual-modality imaging. Biomaterials 32: 6155–6163.
  16. 16. Yang F, Zhang M, He W, Chen P, Cai X, et al. (2011) Controlled release of Fe3O4 nanoparticles in encapsulated microbubbles to tumor cells via sonoporation and associated cellular bioeffects. Small 7: 902–910.
  17. 17. Kaur R, Morris R, Bencsik M, Vangala A, Rades T, et al. (2009) Development of a novel magnetic resonance imaging contrast agent for pressure measurements using lipid-coated microbubbles. J Biomed Nanotechnol 5: 707–715.
  18. 18. Yang F, Li Y, Chen Z, Zhang Y, Wu J, et al. (2009) Superparamagnetic iron oxide nanoparticle-embedded encapsulated microbubbles as dual contrast agents of magnetic resonance and ultrasound imaging. Biomaterials 30: 3882–3890.
  19. 19. Yang F, Li L, Li Y, Chen Z, Wu J, et al. (2008) Superparamagnetic nanoparticle-inclusion microbubbles for ultrasound contrast agents. Phys Med Biol 53: 6129–6141.
  20. 20. Quaia E (2007) Microbubble ultrasound contrast agents: an update. Eur Radiol 17: 1995–2008.
  21. 21. Cai W, Chen X (2007) Nanoplatforms for targeted molecular imaging in living subjects. Small 3: 1840–1854.
  22. 22. Yankeelov TE, Niermann KJ, Huamani J, Kim DW, Quarles CC, et al. (2006) Correlation between estimates of tumor perfusion from microbubble contrast-enhanced sonography and dynamic contrast-enhanced magnetic resonance imaging. J Ultrasound Med 25: 487–497.