Human body acceleration is often used as an indicator of daily physical activity in epidemiological research. Raw acceleration signals contain three basic components: movement, gravity, and noise. Separation of these becomes increasingly difficult during rotational movements. We aimed to evaluate five different methods (metrics) of processing acceleration signals on their ability to remove the gravitational component of acceleration during standardised mechanical movements and the implications for human daily physical activity assessment.
An industrial robot rotated accelerometers in the vertical plane. Radius, frequency, and angular range of motion were systematically varied. Three metrics (Euclidian norm minus one [ENMO], Euclidian norm of the high-pass filtered signals [HFEN], and HFEN plus Euclidean norm of low-pass filtered signals minus 1 g [HFEN+]) were derived for each experimental condition and compared against the reference acceleration (forward kinematics) of the robot arm. We then compared metrics derived from human acceleration signals from the wrist and hip in 97 adults (22–65 yr), and wrist in 63 women (20–35 yr) in whom daily activity-related energy expenditure (PAEE) was available.
In the robot experiment, HFEN+ had lowest error during (vertical plane) rotations at an oscillating frequency higher than the filter cut-off frequency while for lower frequencies ENMO performed better. In the human experiments, metrics HFEN and ENMO on hip were most discrepant (within- and between-individual explained variance of 0.90 and 0.46, respectively). ENMO, HFEN and HFEN+ explained 34%, 30% and 36% of the variance in daily PAEE, respectively, compared to 26% for a metric which did not attempt to remove the gravitational component (metric EN).
In conclusion, none of the metrics as evaluated systematically outperformed all other metrics across a wide range of standardised kinematic conditions. However, choice of metric explains different degrees of variance in daily human physical activity.
The assessment of human daily physical activity in population studies requires accurate, cheap, and feasible measurement technology
Accelerometer devices, based on acceleration sensors which allow for raw data storage expressed in g-units or SI units at a relatively high sampling frequency have been used in gait analysis
A measured acceleration signal consists of a gravitational component, a movement component, and noise
The first two studies that identified the challenge of separating the components of acceleration lacked a comparison against a reference method
The use of gyroscopes in addition to acceleration sensors could be regarded as the solution for separating the gravitational component from the acceleration signal
The main objective of the present study was therefore to evaluate the ability of different methods (metrics) of processing acceleration signals to remove the gravitational component of acceleration by comparison against a reference method under a range of standardised kinematic conditions. A second objective was to assess the shared variance between these metrics in human physical activity data collected during daily life and the impact of metric selection on the accuracy with which daily energy expenditure can be estimated.
Ethical approvals were obtained from the Cambridgeshire research ethics committee, Cambridge (UK) and from the Regional Ethical Review Board in Umeå (Sweden).
The main experiment in this study was done with a robot and did not involve testing of human participants. Two additional sets of experiments were performed, the first to test the degree to which metrics convey similar information when applied to wrist and hip signals, and the second to assess the implication of such differences for estimation of daily physical activity-related energy expenditure.
An industrial robot (TX90, Stäubli Tec-Systems GmbH, Bayreuth, Germany; see
A bar (B) holds five accelerometers and rotates around robot joint (A).
The frequency of oscillation, the radius of rotational movement (shortest distance to centre of rotation), and the angular range of motion were systematically varied. The range of frequency conditions was limited by the maximal amount of mass moment of inertia and torques that could be absorbed by the robot and supporting frame. For all frequencies ranging from 0.05 Hz to 1.2 Hz, eighteen tri-axial accelerometers were positioned along the length of a 70 cm bar mounted to the flange of the robot at 10 cm from the centre of rotation. The application of eighteen accelerometers in parallel allowed for assessment of the relationship between metric output and the radius of movement. To reduce mass moment of inertia at the higher frequencies of oscillation (>1.1 Hz) a shorter bar (20 cm) was used, see
Frequencies | Angle range* | Number of accelerometers (range in position relative to axis of rotation) |
0 Hz | 0° and 22.5° | 18 (0.13–0.78 m) |
0.05 to 0.55 Hz (steps of 0.05) | 0–90° | 18 (0.13–0.78 m) |
0.60, 0.70, and 0.80 Hz | 0–45° | 18 (0.13–0.78 m) |
0.90, 1.00, and 1.10 Hz | 0–20° | 18 (0.13–0.78 m) |
1.20 and 1.30 Hz | 0–45° | 5 (0.13–0.29 m) |
1.4 to 2.6 (steps of 0.1), 2.8, 3.0, 3.2, 3.6 and 4 Hz | 0–20° | 5 (0.13–0.29 m) |
[*for 0° the bar is in horizontal position and for 90° the bar is pointing upwards relative to the axis of rotation].
In order to facilitate the interpretation of the robot experiment in the context of human daily (free-living) physical activity, we asked 47 men and 50 women (healthy, aged 22–65 yrs) to wear accelerometers on their wrist and on their hip for seven days during free-living as previously described
The accelerometer comprised a tri-axial STMicroelectronics accelerometer (LIS3LV02DL) with a dynamic range of ±6 g (1 g = 9.81 m·s−2), as described elsewhere
For the robot analyses three metrics for the estimation of acceleration related to movement were evaluated: (i) the Euclidean norm (vector magnitude) of the three raw signals minus 1, referred to as ENMO; (ii) the application of a high-pass frequency filter (4th order Butterworth filter with ω0 = 0.2 Hz) to each raw signal, after which the Euclidean norm was taken from the three resulting signals,
The third metric has not been described previously. The motivation for metric HFEN+ is as follows: In the absence of rotational movement the Euclidian norm of the three low-pass filtered raw signals (LFEN) is equal to 1 g. In the presence of rotation, however, LFEN may be different to 1 g due to imperfect separation; there we add this difference (positive or negative) to HFEN. A low frequency component above 1 g may result from low-frequency accelerations perpendicular to the direction of rotation, e.g. the centripetal force when sitting on a swing. A low frequency component below 1 g could indicate that part of the gravitational component is still contained in the high-frequency content, e.g. rotations in the vertical plane as a result of which gravity is an alternating component in the signal. A further elaboration on the motivation for metric HFEN+ can be found in
For some of the metrics described above the output could in theory be negative. To gain insight into when this happens, negative values were not corrected for the robot experiment. However, for the accelerometer data collected in daily human movement, negative metric output was rounded off to zero before further analysis.
The filter cut-off frequency of 0.2 Hz for metrics HFEN and HFEN+ was chosen on the presumption that most of daily acceleration related to movement for most human body parts occurs at frequencies higher than 0.2 Hz. n the robot experiment, the exact absolute value of this filter cut-off frequency (0.2 Hz) was considered of minor relevance as this experiment intends to investigate frequency of rotation and frequency of filtering on a relative scale. For the human part of our study, both a cut-off frequency of 0.2 Hz and 0.5 Hz were evaluated to assess the effect of threshold selection in relation to human movement. Additionally, the human part of our study was extended with the application of a band-pass frequency filter version of HFEN (4th order Butterworth filter with ω0 = 0.2–15 Hz), referred to as BFEN, to assess the effect of high-frequency noise removal.
Finally, the Euclidean norm of the three raw acceleration signals (EN) without subtraction of gravity was added to the evaluations in human data to assess the relevance of attempting to remove the gravitational component from an applied perspective.
To sum up, metrics evaluated in this investigation include Euclidian Norm (EN), Euclidian Norm Minus One (ENMO), Bandpass-Filtered followed by Euclidian Norm (BFEN), Highpass-Filtered followed by Euclidian Norm (HFEN), and Highpass-Filtered followed by Euclidian Norm Plus difference between 1 g and low-pass-filtered component (HFEN+).
Reference values for robot acceleration were calculated based on forward kinematics of the robot arm using the radius length (
The average metric output and reference values were calculated over an integer number of oscillating periods in the middle two minutes of each experimental condition (3 minutes), after which absolute and relative measurement errors were expressed. Relative errors were calculated as (Estimated – Reference)/Reference.
For reference purposes, all analyses were repeated based on simulated acceleration signals using the equations as in equation II and equation III. Here,
Metrics ENMO, HFEN, HFEN+, BFEN and EN were applied to the raw data collected on the wrist and hip (7 days) after which metric output was averaged over consecutive non-overlapping 1 minute time windows. Further, metrics ENMO, HFEN, HFEN+, BFEN and EN were applied to the raw data collected in the human participants where PAEE reference data was available. Here, metric output was averaged per person. A detailed description of the detection of monitor non-wear periods and signal clipping are provided in
Means and (relative) differences were computed for the data resulting from the robot experiment. In order to evaluate whether differences between metrics resulted in different measures of free-living human movement, repeated measures ANOVA was used to assess the within- and between-individual explained variance between metrics, stratified by wrist and hip placement. Analyses were performed for all data points excluding non-wear time segments and repeated including imputed data for non-wear time segments. The most important difference is that this would either include or exclude hip accelerometer values for sleeping hours. Results were very similar, and we only report results excluding non-wear time for these analyses. Average and standard deviation of metric output are reported based on imputed data to facilitate the comparison between this study population with future study populations.
For the PAEE analyses, participant inclusion criteria were identical to our previous work
Robot conditions and corresponding reference acceleration are presented in
Metrics | ||||
Frequency (Hz) | Angle (°) | ENMO | HFEN | HFEN+ |
0.05–0.2 | 0–90 | −3.9 | 13.4 | 9.4 |
76.0% | 7.1% | 6.3% | ||
0.25–0.55 | 0–90 | −4.9 | 14.2 | 9.2 |
−12.3% | 2.2% | 2.4% | ||
0.6–0.8 | 0–45 | −2.9 | 18.9 | 15.7 |
−8.1% | 3.7% | 3.7% | ||
0.9–1.1 | 0–20 | 0.9 | 21.5 | 22.0 |
6.7% | 5.8% | 6.4% | ||
1.2–1.3 | 0–45 | 1.5 | 9.3 | 10.8 |
2.0% | 1.4% | 1.9% | ||
1.4–2.0 | 0–20 | 0.1 | 35.9 | 35.2 |
0.4% | 7.8% | 7.9% | ||
2.1–3.0 | 0–20 | 1.4 | 17.1 | 18.3 |
0.7% | 1.9% | 2.1% | ||
3.2–4.0 | 0–20 | 2.3 | 74.8 | 73.5 |
0.2% | 4.3% | 4.3% | ||
0.7 | 25.6 | 24.3 | ||
8.2% | 4.3% | 4.4% |
Relative values are expressed as percentage of average metric output for all accelerometers attached to the bar as fixed to the flange.
The metric output for each accelerometer attached to the bar was compared against the reference acceleration. Metric HFEN+ was more accurate compared to metric HFEN with an average difference in absolute measurement error of respectively, 90 mg and 109 mg. Measurement error was lowest for metric HFEN+ in all but one experimental conditions based on oscillation frequencies higher than 0.2 Hz. On the contrary, metric ENMO outperformed the other metrics for frequencies of oscillation below 0.2 Hz, see
Freq.(Hz) | Angle (°) | Radius (m) | Acc. (mg) | ENMO | HFEN | HFEN+ |
0* | 0 | 0.1–0.3 | 0 | −9 | 4 | −5 |
0* | 0 | 0.3–0.6 | 0 | 0 | 6 | 6 |
0* | 0 | 0.6–0.8 | 0 | −3 | 9 | 6 |
0* | 22.5 | 0.1–0.3 | 0 | −4 | 3 | 0 |
0* | 22.5 | 0.3–0.6 | 0 | −11 | 5 | −4 |
0* | 22.5 | 0.6–0.8 | 0 | −11 | 7 | −4 |
0.05–0.2 | 0–90 | 0.1–0.3 | 14 | −16(−173) | 167 (1427) | 132 (1184) |
0.05–0.2 | 0–90 | 0.3–0.6 | 31 | −38 (−162) | 155 (619) | 112 (447) |
0.05–0.2 | 0–90 | 0.6–0.8 | 48 | −55 (−144) | 152 (442) | 107 (343) |
0.25–0.55 | 0–90 | 0.1–0.3 | 129 | −122 (−98) | 435 (498) | 212 (272) |
0.25–0.55 | 0–90 | 0.3–0.6 | 281 | −251 (−93) | 364 (194) | 89 (76) |
0.25–0.55 | 0–90 | 0.6–0.8 | 434 | −354 (−86) | 308 (108) | −3 (24) |
0.6–0.8 | 0–45 | 0.1–0.3 | 161 | −153 (−97) | 206 (149) | 141 (102) |
0.6–0.8 | 0–45 | 0.3–0.6 | 351 | −328 (−95) | 152 (49) | 57 (21) |
0.6–0.8 | 0–45 | 0.6–0.8 | 541 | −465 (−87) | 118 (24) | 9 (3) |
0.9–1.1 | 0–20 | 0.1–0.3 | 134 | −128 (−99) | 93 (78) | 83 (67) |
0.9–1.1 | 0–20 | 0.3–0.6 | 293 | −292(−100) | 73 (27) | 44 (17) |
0.9–1.1 | 0–20 | 0.6–0.8 | 451 | −419 (−93) | 68 (16) | 35 (8) |
1.2–1.3 | 0–45 | 0.1–0.3 | 508 | −432 (−87) | 160 (35) | 63 (14) |
1.4–2.0 | 0–20 | 0.1–0.3 | 390 | −364 (−95) | 72 (22) | 54 (16) |
2.1–3.0 | 0–20 | 0.1–0.3 | 832 | −618 (−79) | 47 (7) | 22 (3) |
3.2–4.0 | 0–20 | 0.1–0.3 | 1700 | −779 (−50) | 45 (3) | 14 (1) |
Values are average absolute differences in mg (average relative error % in brackets §) between each metric output and the actual acceleration related to movement for various sections of the experiment.
[Acc, average reference acceleration; *zero movement condition; § Relative measurement error was calculated per experimental condition and then averaged across each section of the experiment].
Replication of the analyses with simulated acceleration signals confirmed the empirical findings as described above. A detailed overview of the results based on simulated acceleration signals are included in
When metrics were applied to human wrist and hip acceleration signals collected during free-living conditions, repeated measures ANOVA showed that the shared within- and between-individual variances (r-squared) varied between metric pairs and body locations, see
ω0 (Hz) | EN | ENMO | BFEN | HFEN | HFEN | HFEN+ | HFEN+ | |
0.91 | 0.61 | 0.62 | 0.71 | 0.75 | 0.80 | |||
0.92 | 0.80 | 0.81 | 0.89 | 0.91 | 0.95 | |||
0.58 | 0.80 | 0.99 | 0.96 | 0.96 | 0.93 | |||
0.60 | 0.82 | 1.00 | 0.98 | 0.97 | 0.94 | |||
0.64 | 0.88 | 0.98 | 0.99 | 0.98 | 0.98 | |||
0.74 | 0.91 | 0.97 | 0.97 | 0.98 | 0.99 | |||
0.77 | 0.95 | 0.94 | 0.95 | 0.98 | 0.99 | |||
1016(9) | 32(10) | 114(25) | 118(26) | 93(22) | 110(25) | 94(23) |
[ω0: cut-off for frequency filter].
ω0 (Hz) | EN | ENMO | BFEN | HFEN | HFEN | HFEN+ | HFEN+ | |
ω0 (Hz) | − | − | ||||||
EN | − | − | 0.77 | 0.54 | 0.55 | 0.58 | 0.61 | 0.63 |
ENMO | − | 0.75 | − | 0.89 | 0.90 | 0.92 | 0.94 | 0.95 |
BFEN | 0.11 | 0.46 | − | 1.00 | 0.99 | 0.99 | 0.98 | |
HFEN | 0.10 | 0.46 | 1.00 | − | 0.99 | 0.99 | 0.98 | |
HFEN | 0.11 | 0.48 | 0.98 | 0.98 | − | 0.97 | 0.99 | |
HFEN+ | 0.52 | 0.85 | 0.78 | 0.78 | 0.75 | − | 0.99 | |
HFEN+ | 0.54 | 0.89 | 0.76 | 0.75 | 0.76 | 0.99 | − | |
1007(15) | 18(16) | 46(15) | 48(15) | 42(14) | 50(21) | 45(20) |
[ω0: cut-off for frequency filter].
For the modelling of PAEE, HFEN+ outperformed metrics ENMO, HFEN, BFEN and EN, explaining 36% of the variance in daily PAEE, see
Model input | ω0 (Hz) | SE | R2 | Equation |
EN | − | 0.99 | 0.26 |
−56.146 + BW × 0.023 + EN × 57.093 |
ENMO | − | 0.94 | 0.34 |
−0.172 + BW × 0.025 + ENMO × 0.057 |
BFEN | 0.2–15 | 0.97 | 0.30 |
−0.913 + BW × 0.021 + BFEN × 0.023 |
HFEN | 0.2 | 0.97 | 0.30 |
−0.905 + BW × 0.021 + HFEN × 0.023 |
HFEN | 0.5 | 0.95 | 0.32 |
−0.769 + BW × 0.022 + HFEN × 0.027 |
HFEN+ | 0.2 | 0.93 | 0.36 |
−1.114 + BW × 0.023 + HFEN+ × 0.025 |
HFEN+ | 0.5 | 0.93 | 0.36 |
−0.805 +BW × 0.023 + HFEN+ × 0.026 |
[SE: Residual standard error;
: p<.001;
: p<.01; ω0: cut-off for frequency filter; BW = body weight (kg)].
The present study demonstrates that the choice of signal processing technique for summarising accelerometer data can have a substantial impact on the accuracy with which acceleration related to movement is measured. Subsequently, the choice of signal processing technique impacts on the summary measures of human acceleration data and criterion-related validity for estimating daily PAEE. In the past, physical activity researchers did not have the opportunity to select a metric; the metric decision was made by the manufacturer of the accelerometer
The first and main part of this paper evaluated metrics under a range of standardised kinematic conditions in order to gain insight into how the accuracy of metric output relates to the kinematics of movement. No single metric outperformed all other metrics for all experimental conditions. Metric HFEN+ resulted in less measurement error compared to metric HFEN. This result may indicate that HFEN+ manages to retrieve some of the non-gravitational acceleration in the lower frequency range and/or remove gravitational acceleration from the frequency range above the filter threshold in contrast to metric HFEN. Metric HFEN+ outperformed metrics ENMO and HFEN for the experimental conditions based on oscillating frequencies higher than the cut-off frequency as used by its frequency filter (0.2 Hz), while the ENMO metric outperformed metrics HFEN and HFEN+ for experimental conditions based on oscillating frequencies below this cut-off frequency. This difference between HFEN, HFEN+ and ENMO may partly be explained by the fact that metrics HFEN and HFEN+ aim to remove the gravitational component by making assumptions on its representation in the frequency content of an acceleration signal, while ENMO aims to remove the gravitational component based on assumptions with regard to its magnitude. Metric HFEN+ could be seen as a hybrid version of the two approaches as it relies on both an assumption about the representation of gravity in the frequency domain and an assumption about the magnitude of gravity. The mutual assumption by metrics ENMO and HFEN+ that gravity is measured as 1 g would not hold true if acceleration sensors are not accurately calibrated and would therefore result in biased metric output. Further, metric ENMO has one additional limitation: For a signal with an offset of 1 g (e.g. containing the gravitational component) and an amplitude of less than 1, taking the square will increase the amplitude. On the contrary, if the square is taken from a signal with no offset (e.g. no gravity) and the amplitude is less than one, then taking the square will decrease the amplitude. Therefore, taking the square of three orthogonal signals like in metric ENMO will result in a stronger contribution of vertical accelerations that alternate around 1 g to the resulting summary measure compared with horizontal accelerations that alternated around 0 g.
The reference acceleration as used for the evaluation of the metrics may not have been exactly equal to the true acceleration that the accelerometers were exposed to; imprecision in accelerometer positioning and system vibrations are possible sources of error. In theory, the acceleration of a rotating and non-translating object is proportional to the distance from its centre of rotation, the radius length. A discrepancy of 5 mm (plausible) in the assessment of accelerometer position would represent 0.6% for the accelerometer farthest away and 3.7% for the accelerometer closest to the axis of rotation. This would translate into a similar degree of error in the calculated reference acceleration (0.6–3.7%). Secondly, vibrations of the whole robot during operation may have resulted in the true acceleration exposure being higher than what we calculated it to be. The accelerometer attached to the base of joint 5 did record acceleration beyond the sensor’s noise level likely resulting from the movement of the robot system itself. We believe that robot movement was caused by the supporting frame that vibrated towards the extreme experimental conditions; the robot itself has a high stiffness. The accelerometers attached to the bar mounted on the flange have been exposed to these vibrations as well as those intended by the experimental design. The replication of the robot analyses with simulated acceleration signals confirmed the empirical findings, indicating that environmental vibrations had no significant impact.
As for the analyses conducted on data collected during human daily life, the shared within-individual variances were all above 80% between metrics which make some attempt at removing the gravitational component, indicating the pattern within an individual is picked up quite similarly between those metrics. The between-individual shared variances, which is a measure of the metrics’ ability to rank individuals similarly, showed some differences between hip and wrist positions, most notably lower similarity between ENMO and frequency-filtered metrics for hip than wrist. Whether this reflects differences in monitoring protocols (24-hr vs. non-sleep time), differences in signal to error ratio and/or differences in frequency characteristics of the gravitational component as measured by triaxial accelerometry at these two positions is difficult to conclude from our data. However, it should be noted that shared variances only indicate to what extent metrics are similar in describing variance on a relative level but not what the shared variance represents; it will also include any correlated measurement error and should therefore be interpreted with caution.
Physical activity-related energy expenditure and body acceleration are only distally related to each other. As a consequence, differences in explained variance in daily PAEE does not serve as direct evidence for a metric’s ability to remove the gravitational component.
HFEN+ outperformed HFEN when using daily PAEE as a reference, which confirms the findings from the higher frequency conditions in the robot experiment. Further, ENMO turned out to be a good alternative for HFEN+. The correspondence between the strong performance of ENMO in explaining variance in PAEE in the current analysis with the strong performance of ENMO in the lower frequency range of the robot experiment might indicate that wrist acceleration in daily life is dominated by translational accelerations and/or accelerations resulting from low frequency rotations. A second explanation for the strong performance of metric ENMO may be its higher sensitivity to vertical accelerations (vertical acceleration is amplified) as explained above. The latter would indicate that vertical wrist accelerations are the stronger determinant of daily PAEE compared with accelerations in the horizontal plane. A third and final explanation could be that ENMO is more accurate at measuring translational acceleration compared with some of the other metrics, as the signal is never deformed by frequency filtering in ENMO.
The subtraction of one in ENMO has a constant effect on all the metric output and would in theory be perfectly correlated with EN, which should therefore correlate the same with PAEE. However, there is one additional difference between the two metrics, namely the replacement of negative values by zero in ENMO, which explains why metric ENMO outperforms metric EN for the prediction of PAEE. The truncation of negative values to zero could be hypothesized to be an effective correction mechanism for errors in the subtraction of the gravitational component.
Filter settings for HFEN and HFEN+ were briefly evaluated indicating that a 0.5 Hz filter cut-off frequency may perform slightly better than a 0.2 Hz filter cut-off frequency for predicting PAEE. A more thorough optimization of filter settings could lead to further improvement but also introduces the risk of over-fitting filter configurations to one study population, which may not generalise to others.
One previous study investigated the need for removing the gravitational component using metabolic energy expenditure as reference method and concluded that attempting to remove the gravitational component is not worth the effort
Additional research is needed to explore the potential of combining metrics in a fashion that the best metric is chosen depending on the kinematic conditions. It should be noted that all PAEE-related results apply to the wrist placement and cannot be generalized to other body locations. Future research is therefore also needed to explore the importance of metric selection for other body locations, in particular commonly used positions at the lower back and the hip.
In conclusion, none of the metrics as evaluated systematically outperformed all other metrics across a wide range of standardised kinematic conditions. However, choice of metric explains different degrees of variance in daily physical activity.